• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL FAST-CATH¿ GUIDING INTRODUCER SWARTZ¿ SR SERIES, SR0¿ 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL FAST-CATH¿ GUIDING INTRODUCER SWARTZ¿ SR SERIES, SR0¿ 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number 406853
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
During the svt procedure, it was noted that the sheath was punctured after being in the patient.The sheath was replaced, and the procedure continued without adverse consequences to the patient.
 
Additional Manufacturer Narrative
Additional information: d9, g3, g6, h2, h3, h6, h11.One 8.5f fast-cath introducer sheath and dilator were received for evaluation.The sheath and dilator had been perforated proximal to the distal tip.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ the cause of the perforation is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ GUIDING INTRODUCER SWARTZ¿ SR SERIES, SR0¿ 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
wendy mccarty
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key22664391
Report Number3005334138-2025-00463
Device Sequence Number3716087
Product Code DYB
UDI-Device Identifier05414734202985
UDI-Public(01)05414734202985(10)10381488(17)270531
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406853
Device Catalogue Number406853
Device Lot Number10381488
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2025
Supplement Date Received by Manufacturer09/08/2025
Initial Report FDA Received Date07/30/2025
Supplement Report FDA Received Date09/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-