• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 23CM; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 23CM; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HAR23
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) date sent: 7/30/2025 d4 batch #: u95g7c.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the black coating of the blade damaged.The device was connected to a test handpiece and tested on a gen11.The device did activate during functional testing.The device was analyzed, and it was determined that it was possibly used in more than one procedure, based on generator data.The device is intended and labeled for single patient use.If the device is connected to multiple generators, an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.The device was disassembled to verify the condition of the internal components and no anomalies were found.Occasionally, damage to the blade coating occurs when the blade is exposed to metal contact and/or high heat/prolonged activation without tissue present in the distal section of the jaws and the jaws closed.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.Due to the multiple generator usage, which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and, therefore, cannot conclude the root cause.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch u95g7c, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an unknown procedure the "no instrument uses remaining" screen was displayed by the generator.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 23CM
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key22664436
Report Number3005075853-2025-06009
Device Sequence Number3702358
Product Code LFL
UDI-Device Identifier10705036014430
UDI-Public(01)10705036014430(17)250930(10)U95G7C
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K120729
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue NumberHAR23
Device Lot NumberU95G7C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2025
Initial Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR AND HANDPIECE
Patient SexUnknown
-
-