BECTON DICKINSON CARIBE LTD. BD AFFIRM¿ VPIII MICROBIAL IDENTIFICATION TEST; DNA PROBE, YEAST Back to Search Results
Catalog Number 446257
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd affirm¿ vpiii microbial identification test, there was a false positive result for gardnerella vaginalis.There was no report of patient impact.
 
Additional Manufacturer Narrative
"investigation summary: quality initiated an investigation on customer complaint of affirm gv false positive results.Batch history record review was performed on finished product with satisfactory results.In process and qc testing were within specifications.Functional and specificity tests were performed on retention samples with satisfactory results.This complaint was unable to be confirmed for gv false positive results.Bd was unable to identify a root cause for indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends." this mdr is being submitted as part of a retrospective review of records dating april 2023-2025, under capa 11910483.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD AFFIRM¿ VPIII MICROBIAL IDENTIFICATION TEST
Common Device Name
DNA PROBE, YEAST
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key22664724
Report Number2647876-2025-00239
Device Sequence Number3677786
Product Code MLA
UDI-Device Identifier00382904462574
UDI-Public(01)00382904462574(11)231213(17)241006(10)3347514
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K931374
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/06/2024
Device Catalogue Number446257
Device Lot Number3347514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2024
Initial Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-