| Model Number |
8253200 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that before operation the patient interface box had no stimulation response.During nerve stimulation attempts, no response was detected on the monitor, the stim return showed "0," no "stimulus delivery" tone was heard, and no waveform was displayed.The procedure was completed with the other equipment.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis couldn't be able to verify the reported issue and no fault was observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|