| Model Number |
0174 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a retrospective review of device data it was identified that this right ventricular lead exhibited high out of range shock impedance measurements.No other information was provided.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a retrospective review of device data it was identified that this right ventricular lead exhibited high out of range shock impedance measurements.No other information was provided.No adverse patient effects were reported.Additional information was received that this right ventricular (rv) lead was surgically abandoned due to a product performance issue as stated previously.A new rv lead was implanted and remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.
|
| |
|
Event or Problem Description
|
|
It was reported that during a retrospective review of device data it was identified that this right ventricular lead exhibited high out of range shock impedance measurements.No other information was provided.No adverse patient effects were reported.Additional information was received that this right ventricular (rv) lead was surgically abandoned due to a product performance issue as stated previously.A new rv lead was implanted and remains in service.No additional adverse patient effects were reported.
|
| |
|