MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number SL-2010M2096
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: with this lot number, expires 4/7/28 the arterial line is getting a lot of air as soon as it is hooked up to the fistula or cvc (central venous line).There are air bubbles at that connection even though the user is making sure to lock the device in.The connection is not joining properly.As the devices were attached to a patient - none of the failing product is available for evaluation.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No samples were provided for evaluation.Although no samples were provided, the reported event is similar to a known issue of air in line during use.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.The reported issue is due to a manufacturing issue with the arterial and venous patient connectors, which may allow air leakage resulting in microbubbles in the line, which could thus trigger air-in-line alarms.An approved project is in place to further address issues with air in line.Additionally, b.Braun medical inc.(bbmi) issued a voluntary urgent field safety correction for streamline® bloodline set for dialog+® hemodialysis system (2521402-9/3/25-004-c [z-0070-2026]).We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th avenue
bethlehem, PA 18018
8334251464
MDR Report Key22666368
Report Number2521402-2025-00266
Device Sequence Number19758504
Product Code FJK
UDI-Device Identifier04046964200670
UDI-Public(01)04046964200670(17)280407(10)A2500204
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080807
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSL-2010M2096
Device Lot NumberA2500204
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/30/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0070-2026
Patient Sequence Number1
Patient SexUnknown
-
-