BIOTRONIK SE & CO. KG ACTICOR 7 HF-T DF4 IS-1 PROMRI; CRT-D Back to Search Results
Model Number 429523
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Event or Problem Description
This system was explanted due to infection.Should additional information become available, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTICOR 7 HF-T DF4 IS-1 PROMRI
Common Device Name
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22666427
Report Number1028232-2025-03573
Device Sequence Number17063381
Product Code NIK
UDI-Device Identifier04035479156688
UDI-Public(01)04035479156688(17)251031
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date10/31/2025
Device Model Number429523
Device Catalogue Number429523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2025
Initial Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age57 YR
Patient SexMale
-
-