Following analysis of the device logs, the unit was requested for return and further evaluation.Upon return, the device underwent bench testing, where it was determined that the aed was non-functional and would not have delivered therapy if needed.Further investigation identified the cause as a shorted transistor.Although the original customer report did not involve patient use and was not considered reportable, evaluation confirmed a malfunction that could delay or prevent therapy delivery in a clinical situation.As such, this event is being reported in accordance with 21 cfr 803.50.
|