The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot n178 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot n178 shows no trends.Trends were reviewed for complaint category, pressure dome leak.No trends were detected for this complaint category.Photographs were returned for evaluation.The complaint kit and smart card were not returned for evaluation.Review of the customer provided photographs confirms the report of a pressure dome leak as blood is present at the base of the system pressure dome and on the pump deck.A material trace of the pressure dome housing assembly and its components used to build lot n178 did not find any non-conformances.The device history record (dhr) review did not result in any related non-conformances.This lot passed all lot release testing.The reported pressure dome leak is verified based on the photographs provided; however, the root cause for the pressure dome leak could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4), (b)(6) 2025.
|