COVIDIEN AG JUAREZ (MMJ) SHILEY; AIRWAY, NASOPHARYNGEAL Back to Search Results
Model Number 8888247015
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection, seven out of ten nasopharyngeal tubes total length were found to be outside the limits of detection based on product specifications.There was no patient involved.
 
Additional Manufacturer Narrative
D10 concomitant products: 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx 8888247015 nasopharyngeal airway 24fr lot: 25b0184jzx.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that during inspection, seven out of ten nasopharyngeal tubes total length were found to be outside the limits of detection based on product specifications.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: d3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
AIRWAY, NASOPHARYNGEAL
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22668242
Report Number9611018-2025-00005
Device Sequence Number15355385
Product Code BTQ
UDI-Device Identifier60884521133070
UDI-Public(01)60884521133070(17)300211(10)25B0184JZX(20)02
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K760125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888247015
Device Catalogue Number8888247015
Device Lot Number25B0184JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer10/09/2025
04/02/2026
Initial Report FDA Received Date07/30/2025
Supplement Report FDA Received Date10/15/2025
04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTES ON H11.
Patient SexUnknown
-
-