| Catalog Number |
1-108 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The physician who performed an omni procedure reported that the catheter severed while performing the 180-degree goniotomy procedure.The severed portion of the catheter was removed using forceps.A partial goniotomy occurred while removing the severed portion of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was carefully evaluated and was determined that there is no evidence that the device design or manufacturing process was the cause of the incident.A review of device lot history and analysis of the most suspected components (i.E., cannula and catheter) concluded that there were no production non-conformances or any other anomalies that would have contributed to this event.Engineering had previously recreated the severing of catheter failure which did not point to a design or manufacturing issue as the cause of the failure.The mode of failure does suggest it's associated with a use error.The ifu provides cautionary statements related to possible kinking or damaging (e.G., severing) of microcatheter should the steps not be correctly performed.Therefore, it is concluded that the probable root cause for this event is likely due to a surgical technique or use error.
|
| |
|