• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3507300BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Deformity/ Disfigurement (2360)
Date of Event 07/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a breast surgery with implantation of a 300cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with bilaterally ruptured breast implants using an undisclosed imaging methodology.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.Follow-ups have been conducted, and no additional information has been received.If more information becomes available, mentor will submit a supplemental report accordingly.
 
Additional Manufacturer Narrative
On august 25, 2025, a manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On september 18, 2025, mentor was notified that the patient was also diagnosed with bilateral baker grade iv capsular contracture.Additionally, bilateral implant exchange was performed on (b)(6) 2025.Date of explantation: (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: rupture, capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On october 21, 2025, the mentor analysis lab received the suspect medical device for evaluation.On november 14, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted visual inspection of the device.Visual analysis of the returned device revealed that the breast implant was ruptured from the anterior aspect expanding to the posterior aspect.Additionally, it had an area of shell abrasion on the edges.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22669066
Report Number1645337-2025-08353
Device Sequence Number19752277
Product Code FTR
UDI-Device Identifier00081317000099
UDI-Public(01)00081317000099(17)140828(10)5929900
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2014
Device Catalogue Number3507300BC
Device Lot Number5929900
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer08/25/2025
09/18/2025
10/21/2025
Initial Report FDA Received Date07/30/2025
Supplement Report FDA Received Date08/25/2025
10/03/2025
11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-