• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Catalog Number 71118200
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during internal fixation surgery, while passing the flexible shaft w/cir connector into the canal for the third time, it fractured at the distal end.No broken pieces fell inside the patient.The procedure was completed without any surgical delay using the same device.No injury was reported as a consequence of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Common Device Name
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key22669069
Report Number1020279-2025-01285
Device Sequence Number11681922
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public3596010436528
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K121714
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71118200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Initial Report FDA Received Date07/30/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age31 YR
Patient SexMale
Patient Weight75 KG
-
-