SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 XS/-3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
|
Back to Search Results |
|
| Catalog Number |
75100855 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a thr surgery, the inner tab of the trial femoral head 36 xs/-3 broke while attempting to reduce the head trial into the liner socket.The broken off piece was found and successfully recovered quickly.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation, however, previous complaints were received and investigated for this part number, in which the reported failure mode was confirmed.Therefore, no further investigation tasks are deemed necessary.The root cause of the event can be attributed to a known inherent risk of the device.Previous investigations demonstrated that the trial femoral head may disconnect from the stem taper intra-operatively and flaps might bend or break off.These failure modes may further be exacerbated by excessive use of the device over time.The performance of the device is still within the risks level that are anticipated in the risk management documentation.The current design will be further monitored through post-market surveillance processes.No further actions are necessary at this time point.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4, d9, h3, h4, h6 and h11.H11: the complaint device intended for use in treatment not been completely returned.A visual evaluation of the device was conducted, and it was concluded that one of the inner flaps fractured off and is missing.Furthermore, there are signs of wear such as scratches and dents.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The review of historical complaints for the alleged device revealed 4 additional similar complaints reported for the same batch, and 5 additional similar complaints for the same product number over the past 12 months with similar a failure mode.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The root cause of the event can be attributed to a known inherent risk of the device.Previous investigations demonstrated that the trial femoral head may disconnect from the stem taper intra-operatively and flaps might bend or break off.These failure modes may further be exacerbated by excessive use of the device over time.The performance of the device is still within the risks level that are anticipated in the risk management documentation.The current design will be further monitored through post-market surveillance processes.No further actions are necessary at this time point.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference: (b)(4).
|
| |
|