Model Number H74939401103250
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the balloon was difficult to remove and an additional device was required to remove the balloon.The 95% stenosed target lesion was located in the mildly tortuous and severely calcified right coronary artery.A 10mm x 3.25mm wolverine coronary cutting balloon monorail was selected for use.During the procedure, it was noted that after the fifth time the wolverine was deflated, removal attempt was unsuccessful.The device was difficult to remove from the ostial lesion and difficult to get back into the non-boston scientific guide catheter.A guidezilla was advanced over the balloon to capture and remove it in one piece.The balloon was fully deflated.The procedure was completed with the original device.No further patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22669203
Report Number2124215-2025-51403
Device Sequence Number7607789
Product Code NWX
UDI-Device Identifier08714729888277
UDI-Public(01)08714729888277(17)250710(10)31987368
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/10/2025
Device Model NumberH74939401103250
Device Catalogue NumberH74939401103250
Device Lot Number0031987368
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2025
Initial Report FDA Received Date07/30/2025
Date Device Manufactured07/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-