Corrective data: on block h6, the medical device problem code "3190" was updated to "2993" following the investigation results.Additional manufacturer narrative: (4111) following our surgeon contact, the following information has been provided: the procedure was prolonged by 30 to 40 minutes, there is no consequence for the patient.The pre-operative medications provided to the patient was: platelet inhibitors, anticoagulants, heparin, ascal and clopidogrel (herparin before clamping and all medication as given to this patient is standard procedure) the vascular diagnosis: 7/6 tia bij stenose carotis, the procedure performed was a carotid endarterectomy , indication for surgery: tia , the amount of blood loss was 300cc, the patient did not required blood products, actions taken to stop the bleeding: protamine, tachosil and waiting the hemostasis, where was the patch implanted at the carotid bifurcation, the surgeon is not a new user intergard product.(4112/213) the case and its investigation have been reviewed by the medical affairs department who concluded that: "it is unclear if the bleeding can be attributed to the patient¿s coagulation profile or the patch.Needle hole bleeding is not uncommon and can be caused by surgical technique, altered coagulation profiles, or inherited bleeding disorders.While some bleeding was observed along the surface of the patch, the main concern reported was the needle hole bleeding.Needle hole bleeding would not likely be caused by the patch." (22/4310) as indicated in the medical assessment, needle hole bleeding is not uncommon and can be attributed to the surgical technique, altered coagulation profiles, or inherited bleeding disorders.While some bleeding was observed along the surface of the patch, the main concern reported was the needle hole bleeding.The adverse event would not likely be caused by the patch.The conducted investigation suggests that the product was not defective at the time of manufacturing.It should be noted that bleeding is an undesirable side effect as indicated in the current product instructions for use, in the "undesirable side effects" section.
|