| Model Number |
02.12.E0410FR |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
07/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 22 july 2025.Gmk-sphere 02.12.E0410fr gmk-sphere tibial insert e-cross - flex 4r - 10mm (k202022) lot 2500606: (b)(4) items manufactured and released on 26/02/2025.Expiration date: 11/02/2025.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Conclusions: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|
Event or Problem Description
|
|
The patient came in due to signs of an infection and the pathogen is unknown.About 3 weeks after the primary surgery, the surgeon performed a washout and revised the liner.The surgery was completed successfully.
|
| |
|