| Catalog Number |
TRS-VATS-15 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device, trs-vats-15, anchor tissue retrieval system, was being used during an unknown procedure on an unknown date and when it was reported, ¿bag tore during procedure.¿ follow up questioning found that the device did fragment, and the fragmentation was retrieved from the patient using a grasper.There was no report of injury, medical intervention, or hospitalization of the patient.The procedure was completed with an alternate device.The patient was discharged as normal from the facility.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the device nor photographic evidence was provided.Therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 59 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to not use the anchor¿ tissue retrieval system¿ if resistance is met upon deployment.Improper use of the anchor¿ tissue retrieval system¿ may result in perforation of tissue and subsequent bleeding.Care should be taken when using sharp and electro-surgical devices.Note the size of the trocar cannula when removing a specimen through the trocar cannula and seal system.If required, enlarge the port site to aid removal of the anchor¿ tissue retrieval system¿ by conmed.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device, trs-vats-15, anchor tissue retrieval system, was being used during an unknown procedure on an unknown date and when it was reported, ¿bag tore during procedure.¿ follow up questioning found that the device did fragment, and the fragmentation was retrieved from the patient using a grasper.There was no report of injury, medical intervention, or hospitalization of the patient.The procedure was completed with an alternate device.The patient was discharged as normal from the facility.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|