• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR 3.5MM HOOK 1-PIECE ELECTRODE -EA; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR 3.5MM HOOK 1-PIECE ELECTRODE -EA; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 227305
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported that during an unspecified arthroscopy surgical procedure, when using the vapr 3.5mm hook 1-piece electrode device, it was observed that part of the coating came off and the hook broke.It was reported that the device was new.There were no reported adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary: the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR 3.5MM HOOK 1-PIECE ELECTRODE -EA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22675198
Report Number1221934-2025-02917
Device Sequence Number6026269
Product Code GEI
UDI-Device Identifier10886705009718
UDI-Public(01)10886705009718(10)U2411001(17)271112
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K082643
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227305
Device Lot NumberU2411001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date07/31/2025
Supplement Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-