• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4; BONE PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4; BONE PLATE Back to Search Results
Catalog Number 400.834
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported during a excision of saddle area occupying lesion surgery, when inserting the screw, the screw broke.Surgeon changed to another screw to continue the surgery; the same problem happened 3 additional times.All the pieces were removed from the patient.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H.11.Additional narrative: corrected: g1 (manufacturing site name.H3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the cranial-scr plusdrive ø1.6 self-drill l4 was broken and slightly deformed at the tip.The broken fragment was not returned for evaluation.However, the fracture surface was thoroughly examined on the screw and no problems with the material were noted.The observed condition is consistent with a torsional failure which is likely due to the application of significant torque during the implantation process.According to the low-profile neuro the following warnings are mentioned: these devices can break during use (when subjected to excessive forces or outside the recommended surgical technique).Be aware that implants are not as strong as native bone.Implants subjected to substantial loads may fail.A dimensional inspection could not be performed due to the damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the cranial-scr plusdrive ø1.6 self-drill l4 was consistent with the reported condition.Based on the investigation findings, the potential cause is traced to the application of significant torque during implantation.There is no indication that a design or manufacturing issue has caused the reported complaint condition and it has been determined that no corrective and/or preventative action is proposed.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 400.834.Lot # 39031p8.Manufacturing site: werk selzach logistik.Manufacturing date: 18.Oct.2024.Supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9, h4.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4
Common Device Name
BONE PLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach CO 2425
SZ   2425
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22675201
Report Number8030965-2025-07852
Device Sequence Number15003561
Product Code JEY
UDI-Device Identifier20887587045402
UDI-Public(01)20887587045402(10)39031P8
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K022012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number400.834
Device Lot Number39031P8
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/16/2025
Supplement Date Received by Manufacturer08/08/2025
08/25/2025
Initial Report FDA Received Date07/31/2025
Supplement Report FDA Received Date08/11/2025
08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-