|
Blank fields on this form indicate the information is unknown or unavailable.E1: phone = (b)(6).H3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, during an aortic arch repair procedure, the hub of a flexor raabe guiding sheath separated from the device.The device was used coaxial within an unspecified 10-french sheath, over an unknown 0.035-inch wire guide.During the ¿cross over, coaxial maneuver¿, the hub separated.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures or experience any adverse effects due to this event.Additional information has been requested.
|
|
Summary of event: as reported, during an aortic arch repair procedure, the hub of a flexor raabe guiding sheath separated from the device.The device was used coaxial within an unspecified 10-french sheath, over an unknown 0.035-inch wire guide.During the ¿cross over, coaxial maneuver¿, the hub separated.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures or experience any adverse effects due to this event.Additional information was requested; however, the customer did not respond.Corrected information: d4.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation; however, the customer provided photos, showing that the hub separated from the sheath.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the lot.A review of complaint history found no additional complaints for this lot number.The product ifu warns ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and inspection of photos provided by the customer suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a component failure contributed to this event.The customer did not respond to multiple requests for additional information, and the device was not returned to cook.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|