(b)(4).Date sent: 07/31/2025.D4: batch #unk.The photo upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr45b with lot number 359c88; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4).Date sent: 08/17/2025.D4: batch #292c43.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr45b reload was received with no apparent damage and fully fired.No functional test could be performed due to the condition of the reload.Additionally, photos were provided, one photo shows surgical scene where cut tissue can be observed oozing blood, a second photo and a video showed three green reload and four blue reload, these reloads were fully fired, no apparent damage could be observed from the provided photo and video.The event described could not be confirmed as the reload was received fully fired.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: when positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc.Are within the instrument jaws.Firing over an obstruction may result in incomplete cutting action, improperly formed staples, and/or inability to open the instrument jaws.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 292c43, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|