| Catalog Number |
159547 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Dislocation (2374)
|
| Date of Event |
07/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - associated medical devices: oxf uni tib tray sz c lm pma; item# 154722; lot# 3639874.Oxford uni twin-peg femoral md; item# 166942; lot# j3725800.G2 - foreign: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately eight years and ten months post-implantation, the patient underwent a revision due to bearing dislocation and pain.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.No product was returned but two pictures were provided that show the articulating surface of the bearings.The surface shows scratches and signs of wear and deformation.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination/s.Complaints are monitored per complaint trending process.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|