MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SHERLOCK (QUALITEL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SHERLOCK (QUALITEL) Back to Search Results
Model Number N/A
Medical Device Problem Code Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.
 
Event or Problem Description
It was reported new sensor is producing inconsistent p-waves and is giving negative p-waves, even after pulling back the picc.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.During evaluation, the reported issue of the sensor is producing inconsistent pwaves and is giving negative pwaves even after pulling back the picc was confirmed.When connecting an ecg simulator, it was noticed that the fin assembly was loose when attached, and the yellow p-wave shows up intermittently.The root cause of the reported failure was identified as the ecg pins on the top housing being worn from use of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHERLOCK (QUALITEL)
Common Device Name
SHERLOCK (QUALITEL)
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city UT 84116
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key22676266
Report Number3006260740-2025-05447
Device Sequence Number8892442
Product Code LJS
UDI-Device Identifier00801741163043
UDI-Public(01)00801741163043(11)240417
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K061240
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number9770352
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2025
Supplement Date Received by Manufacturer10/01/2025
Initial Report FDA Received Date07/31/2025
Supplement Report FDA Received Date10/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-