• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS; FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS; FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G34874
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - phone number: (b)(6).H3: device evaluation is anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported the flexor parallel ureteral access sheath and dilators¿ sheath was delaminated during an unspecified procedure.The sheath was internally stripped and generated numerous filaments.The procedure was completed by using same-type device.Additional information has been requested but has not yet been received.
 
Additional Manufacturer Narrative
Investigation evaluation description of event: it was reported the flexor parallel ureteral access sheath and dilators¿ sheath was delaminated during an unspecified procedure.The sheath was internally stripped and generated numerous filaments.The procedure was completed by using same-type device.Reviews of documentation including the complaint history, quality control procedures, manufacturing instructions (mi), and instructions for use (ifu) were conducted during the investigation.The complaint device was not returned.It was assumed the filaments from the inner lining of the sheath.It is possible that the other devices inserted through the scope peeled the liner.No information was known about other devices used in the procedure.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.The ifu did not provide any information related to the reported issue.Cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS
Common Device Name
FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key22676336
Report Number1820334-2025-00921
Device Sequence Number18429206
Product Code FED
UDI-Device Identifier00827002348745
UDI-Public(01)00827002348745(17)280303(10)16490628
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K172217
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG34874
Device Catalogue NumberFUS-107045-P
Device Lot Number16490628
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2025
Supplement Date Received by Manufacturer08/28/2025
Initial Report FDA Received Date07/31/2025
Supplement Report FDA Received Date09/02/2025
Date Device Manufactured03/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-