Investigation evaluation description of event: it was reported the flexor parallel ureteral access sheath and dilators¿ sheath was delaminated during an unspecified procedure.The sheath was internally stripped and generated numerous filaments.The procedure was completed by using same-type device.Reviews of documentation including the complaint history, quality control procedures, manufacturing instructions (mi), and instructions for use (ifu) were conducted during the investigation.The complaint device was not returned.It was assumed the filaments from the inner lining of the sheath.It is possible that the other devices inserted through the scope peeled the liner.No information was known about other devices used in the procedure.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.The ifu did not provide any information related to the reported issue.Cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|