| Catalog Number |
SHPX355 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with mentor memorygel xtra breast implants experienced capsular contracture baker grade iii on the right side post-operatively, which was confirmed during a physical exam.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
On nov 3, 2025, additional information was received indicating the patient experienced bilateral capsular contracture baker grade iii.The explantation surgery is scheduled for (b)(6) 2025.This report is for the right breast prosthesis.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On jan 28, 2026, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|