| Catalog Number |
0168L16 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that while using product nursing staff were placing a foley catheter and the valve for filling the balloon with sterile water came off the line, causing the contents to leak.They were unable to re-attach the valve.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: d, g.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that while using product nursing staff were placing a foley catheter and the valve for filling the balloon with sterile water came off the line, causing the contents to leak.They were unable to re-attach the valve.Per additional information received via mail on 13aug2025, stated that during foley placement for intravesical treatment the valve for filling the balloon with sterile water came off the line, causing the contents to leak.Writer tried to re-attach valve without success, subsequently filled balloon with 10 cc normal saline and clamped with locking hemostat.Catheter placement maintained for treatment without discomfort to patient and provided lot#ngjz0494.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states instructions for use: follow your facility protocol for all processes related to catheterization, re-catheterization, stabilization and urine collection.Follow aseptic procedures outlined by your local hospital or healthcare facility.Thoroughly wash hands, wear sterile gloves, and use aseptic technique whenever handling the catheter or catheter components during insertion.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, e udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that while using product nursing staff were placing a foley catheter and the valve for filling the balloon with sterile water came off the line, causing the contents to leak.They were unable to re-attach the valve.Per additional information received via mail on 13aug2025, stated that during foley placement for intravesical treatment the valve for filling the balloon with sterile water came off the line, causing the contents to leak.Writer tried to re-attach valve without success, subsequently filled balloon with 10 cc normal saline and clamped with locking hemostat.Catheter placement maintained for treatment without discomfort to patient and provided lot#ngjz0494.
|
| |
|