• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SSF-605
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 03/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "infection (unknown onset)" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection (unknown onset).
 
Event or Problem Description
Healthcare professional reported "infection".This record is for the "left" side.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a4, b3, b6.
 
Event or Problem Description
Healthcare professional reported "infection".This record is for the "left" side.The device was explanted.
 
Event or Problem Description
Healthcare professional reported "infection".This record is for the "left" side.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: h11.Further investigation: with the information collected during the investigation, there is enough evidence to support that units involved in this work order were manufactured under controlled conditions in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents related to the reported event.Seal strength was successfully completed and results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run 30052345 sterilization run is not related to any additional complaint infection record reported as of today.Considering that there is no adverse trend for this type of event, no additional actions are deemed required at this time.The issue with infection will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Given the fact that the cause of the infection cannot be specifically associated with the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) were identified during the investigation associated with this lot and the complaint, no corrective action is deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22679658
Report Number9617229-2025-12878
Device Sequence Number21055900
Product Code FTR
UDI-Device Identifier10888628033214
UDI-Public(01)10888628033214(10)1253607(11)240717(17)290617
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSSF-605
Device Lot Number1253607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer07/30/2025
08/19/2025
Initial Report FDA Received Date07/31/2025
Supplement Report FDA Received Date08/18/2025
09/09/2025
Date Device Manufactured07/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
Patient Weight80 KG
-
-