STRYKER ORTHOPAEDICS-MAHWAH MRH TIBIAL B/PLT KEEL MED 2; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
64813112 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
03/13/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following was reported: "aseptic loosening.All components removed.Left knee.".
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening involving a mrh baseplate was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional, and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: could not be performed as the device lot details were not provided.Complaint history review: could not be performed as the device lot details were not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|