W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT231412 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm expansion (4597)
|
| Date of Event |
06/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent an emergency endovascular treatment for an impending rupture of an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprosthesis and gore® excluder®aaa endoprosthesis.The trunk-ipsilateral leg endoprosthesis was deployed below the right renal artery, but the device moved distally about 5 mm due to neck angulation.There was no endoleak, and the procedure was completed.Treatment of the left common iliac artery aneurysm was planned as a staged procedure.On an unknown date, postoperative follow-up examination showed the aneurysm enlargement.On (b)(6) 2025, a reintervention was performed.Although no obvious endoleaks were identified, an additional stent graft was implanted in the proximal side to address a suspected proximal type i endoleak.For a suspected type ii endoleak via the iliolumbar artery, embolization of the lumbar artery was performed.As part of the staged treatment for the left common iliac artery aneurysm, the left internal iliac artery was embolized, and an additional stent graft was placed in the left external iliac artery.
|
| |
|
Search Alerts/Recalls
|
|
|