Additional information provided in d.9., h.2., h.3., h.6.And h.11.Opened sleeves were visually inspected.Visual inspection confirmed the shafts of the infusion sleeves were consistent with damaged observed when the infusion sleeves are pierced by the phaco tip.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is related to customer handling during the surgical set up process.Improper mounting of the infusion sleeve on the handpiece can cause a tear (or entire shaft removal) on the infusion sleeve, leading to the customer's reported experience.No further investigation or manufacturing actions are warranted at this time as user handling is suspected.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|