|
Radiometer reference number: (b)(4) according to the complaint unplausible ph values were measured on (b)(6) 2025 at 13:27 and 13:59; the abl800 flex plus analyzer (serial no; (b)(6) was displaying a green status at this time.Comparative measurements were performed on 3 abl800 devices using 2 patients from different safe pico a syringes (lot dx17), with the ph value on the faulty device being approx.0.2 too low.Operating materials used: cal 1 lot cz03, cal 2 lot cz02, rinsing solution lot dz65, cleaning solution cy02, gas 1 y29b68, gas 2 y17g82, ref.Membrane lot r3594.Sample/patient id (b)(6), 13:37, (b)(6) 2025, ph 7,036 (discrepant), sample/patient id (b)(6), 13:44, (b)(6) 2025, ph 7,202 (comparison), sample/patient id (b)(6), 13:52,(b)(6) 2025, ph 7,207 (comparison).No one was injured.The abl800 flex plus analyzer is running fine after electrode replacement.
|