• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO COVERSTAINER; AUTOMATED SLIDE STAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO COVERSTAINER; AUTOMATED SLIDE STAINER Back to Search Results
Model Number CS100
Medical Device Problem Codes Break (1069); Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an agilent field service engineer was onsite for an unrelated issue and found that the lower door of the coverstainer was cracked.The picture provided by the field service engineer shows a crack at the upper left corner of the lower door confirming the reported issue.The field service engineer serviced the instrument with replacement of the lower door and shock absorbers.The instrument is fully operational and ready for use.The customer was able to complete staining.
 
Additional Manufacturer Narrative
While there was no direct, or indirect, user harm reported for this event; it is being reported due to the potential for harm if the event were to reoccur.The instrument experienced a similar failure mode related to reportable complaint with injury, reference number mdr 3003423869-2023-00084, mhra (b)(4), agilent (b)(4).Therefore, this report is being filed as part of agilent's commitment to due diligence reporting.Agilent technologies initiated a recall april 2024 with the scope of notifying all coverstainer customers of the following issue: the lower door that opens to the reagent bottles is cracking and/or breaking off in the upper corner of the plexiglass; recall event id 94439.Capa01211 has addressed the issue and the investigation has been completed.Agilent technologies has determined the root cause as such; the lower door support bracket, which connects the hinges to the lower door, is only attached to the plexiglass and not to the metal frame which is the main support structure of the lower door.The weight of the plexiglass door is in turn cracking/breaking at this joint.Customers have been notified of this issue via a customer letter.Agilent technologies worked with the manufacturer to re-design the support bracket for the lower door hinges, so the weight of the lower door assembly is not supported by the plexiglass, but by the metal frame of the lower door.This new bracket was installed during this visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAKO COVERSTAINER
Common Device Name
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, sgp 76892 3
SN  768923
Manufacturer Contact
benjamin gaither
1834 state highway 71 west
cedar creek, TX 78612
5123328207
MDR Report Key22687043
Report Number3003423869-2025-00041
Device Sequence Number9472804
Product Code KPA
UDI-Device Identifier05700571112571
UDI-Public(01)05700571112571(11)110929
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Notification,Modification/Adjustment,Repair,Recall
Type of Report Initial
Report Date (Section B) 08/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCS100
Device Catalogue NumberCS10030
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2025
Initial Report FDA Received Date08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1919-2024
Patient Sequence Number1
Patient SexUnknown
-
-