• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 3000TFX
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 07/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a 23mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of seven (7) years, eight (8) months due to severe stenosis.Patient presented with symptoms of heart failure - shortness of breath, chest tightness.The procedure was performed with a 23mm 9755rsl transcatheter valve successfully.Patient was in stable condition post procedure.Patient was discharged on pod #1.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the subject device is not available for evaluation as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections g3, h6 (type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia, diabetes mellitus, atherosclerosis, coronary artery disease.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key22688399
Report Number2015691-2025-06238
Device Sequence Number17645856
Product Code DYE
UDI-Device Identifier00690103163650
UDI-Public(01)00690103163650(17)210827
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/27/2021
Device Model Number3000TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer08/25/2025
Initial Report FDA Received Date08/01/2025
Supplement Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexMale
-
-