• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW REDIGUARD IAB: 7FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW REDIGUARD IAB: 7FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAB-S730C
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# b)(4).
 
Event or Problem Description
It was reported "balloon would not inflate.The balloon was removed, and a new balloon was opened.No safety concerns with patient".No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The serial number on the returned sample is (b)(6).Returned for investigation was a 30cc 7.0fr rediguard intra-aortic balloon catheter (iabc) without the original packaging.The sample was re-turned in the ups shipping box and was in a sealed bio-hazard bag.Returned with the sample was supplied 30cc inflation driveline tubing; no damage or abnormalities were noted to the returned inflation driveline tubing.Upon return, the one-way valve was tethered to the short driveline tubing.The iabc bladder was fully unwrapped.Dried blood was noted on the exterior surfaces of the returned sample.No blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The bladder thickness was measured at six points with measurements ranging from 0.0063in-0.0068in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was connected to an iabp with the returned 30cc inflation driveline tubing.The iabc was inserted into the "t" tube and 75mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for iab "balloon would not inflate" is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.During the investigation, the iabc fully inflated, and no leaks/alarms were noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
 
Event or Problem Description
It was reported "balloon would not inflate.The balloon was removed, and a new balloon was opened.No safety concerns with patient".No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW REDIGUARD IAB: 7FR 30CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key22688451
Report Number3010532612-2025-00709
Device Sequence Number7046657
Product Code DSP
UDI-Device Identifier00801902002686
UDI-Public(01)00801902002686(11)231028(17)250930(10)18F23K0047(21)
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K981660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/30/2025
Device Catalogue NumberIAB-S730C
Device Lot Number18F23K0047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date08/01/2025
Supplement Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexUnknown
-
-