The hospital reported that during an endoscopic vein harvesting procedure, hemopro2 extension cable was observed that cutting tool was not working.They observed that extension cable was defective, they completed the procedure with the new cable.The only troubleshooting step taken was checking with the saline test.There was no delay.There was no patient harm.
|
|
Tw # (b)(4).Updated sections: b4,d9,g3,g6,h2,h3,h4,h6,h11.Corrected section: d4 udi#, exp date.The device was returned to the factory for evaluation on 02/02/2026.An investigation was conducted on 02/05/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on both connector ends.There were no visual defects observed on the intact cable.The cable was able to be connected to the adaptor with no physical or visual difficulties.The reference harvesting device was attached to the cable with no physical or visual difficulties observed.An electrical evaluation was conducted.A pre-cautery test was performed per the instruction for use (ifu) with a reference hemopro 2 evh tool, reference adapter and reference power supply set at level 3.0.The device passed the pre-cautery test.The power supply emitted an audible beeping sound and the evh reference tool produced steam and heat.An activation and transection capability test was performed over four (4) repetitions using "max life test method.The reference device successfully transected tissue four (4) times.Based on the returned condition of the device and the results of the evaluation, the reported failure "failure to deliver energy" was not confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
|