| Model Number |
9797 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed.If the device is returned for evaluation or new information is obtained, a follow up report will be submitted.Health effect impact code: no adverse event, no patient impact.E1.Initial reporter phone number exceeded maximum allowable digits, phone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the device "turns off by itself, for no reason".There was no patient impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned device was evaluated.During evaluation the device passed all visual and functional testing.The device was found to power on and off using both ac and battery power.The device was also found to visually and audibly alarm during alarm conditions.The device was confirmed to be functioning as designed.E1: initial reporter phone number exceeded maximum allowable digits, phone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the device "turns off by itself, for no reason".There was no patient impact or consequence reported.
|
| |
|