Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary - it was brought to the attention by spd that the t-handle plastic is cracked and holding bio burden.The modular endo trial handles are holding bio burden in the ball pieces that the trail head snaps onto.The event did not happened during a surgery.A replacement needs to be shipped.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed signs of heavy usage on the overall device surface.Additionally, some debris was found attached to the ball plungers of the s/c trial handle.No cracks/breaks were identified in the device.No other anomalies were noted.Per ifu-0902-00-721, care should be taken to remove any debris, tissue, or bone fragments that may collect on the instrument.Instruments with cannulas, moving parts, or spring mechanisms require additional cleaning steps to ensure proper removal of all trapped debris.In addition to other important steps to follow, the ifu instructs the use of a soft bristle brush to remove all traces of blood and debris, paying close attention to threads, crevices, seams, and any hard-to-reach areas of the device features.If the instrument has sliding mechanisms or hinged joints, actuate the area to free any trapped blood and debris.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the s/c trial handle would have contributed to a device issue.Based on the investigation findings, the potential cause is traced to maintenance, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot - a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code pg0509 provided is not a valid finished goods lot number.Added: h6 medical device problem code.Corrected: h3.
|