• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_12.6 (-7.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 06/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - work order search: no additional similar complaint type events within associated lots were found.Refractive surprise - each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a shorter length lens.The problem was resolved.Cause of the event was reported as patient related.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.
 
Additional Manufacturer Narrative
Corrected data: b5 - the reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a different power lens.The problem was resolved.Cause of the event was reported as patient related.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key22691851
Report Number2023826-2025-01099
Device Sequence Number8250845
Product Code MTA
UDI-Device Identifier00840311300549
UDI-Public(01)00840311300549(11)240627(17)270531
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberVICM5_12.6 (-7.0)
Device Catalogue NumberVICM5_12.6
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2025
Supplement Date Received by Manufacturer11/09/2025
Initial Report FDA Received Date08/01/2025
Supplement Report FDA Received Date11/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-SFC-45, LOT#-1771289.; CARTRIDGE MODEL-SFC-45, LOT#-UNK.; FOAM TIP PLUNGER MODEL-FTP, LOT#-1727990.; FOAM TIP PLUNGER-MODEL-UNK, LOT#-UNK.; INJECTOR MODEL-MSI-PF, LOT#-158576.; INJECTOR MODEL-MSI-PF,LOT#-UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexFemale
-
-