| Model Number |
VICM5_12.6 (-7.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
06/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no additional similar complaint type events within associated lots were found.Refractive surprise - each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a shorter length lens.The problem was resolved.Cause of the event was reported as patient related.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5 - the reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a different power lens.The problem was resolved.Cause of the event was reported as patient related.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.Claim#: (b)(4).
|
| |
|
Search Alerts/Recalls
|