EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 6900P
Medical Device Problem Code Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446)
Date of Event 06/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11.Additional narrative/data.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a patient with a 27mm 6900p mitral valve underwent a valve-in-vale procedure after an implant duration of seven (7) years and five (5) months due to central regurgitation.The patient presented with heart failure and dyspnea.The procedure was performed with a 29mm 9755rsl transcatheter valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections g3, g6, h2, h6 (type of investigation, investigation findings, investigation conclusions).Regurgitation identified post-procedurally has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Regurgitation occurring post-procedurally is most commonly related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing non-conformance.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including chronic kidney disease and hyperlipidemia.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified regarding warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES, LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22695576
Report Number2015691-2025-06271
Device Sequence Number8250957
Product Code DYE
UDI-Device Identifier00690103152647
UDI-Public(01)00690103152647(17)210816
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/16/2021
Device Model Number6900P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer08/29/2025
Initial Report FDA Received Date08/02/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age65 YR
Patient SexMale
-
-