| Catalog Number |
966120 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty, during the implantation of the sigma cr with the high performance surgical instrument set, a cracking sound was heard during routine use of the rod; after removing the rod from the canal, it was discovered that the distal part remained in the canal.Attempts to extract a fragment of the tip of the rod were made without success.Patient was discharged.Plan for patient management is observation.
|
| |
|
Search Alerts/Recalls
|