| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation revealed the substrate of the device's circuitry to be broken.The investigation results appear to match the problems mentioned in the recipient report.The exact reason why the crack occurred cannot be determined, however a review of the dhr has been performed and no abnormality was revealed, thus the device was released according to specifications.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.There was no redness or swelling seen over the implant area.Re-implantation took place on (b)(6) 2025.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.There was no redness or swelling seen over the implant area.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|