Model Number MI1210 SYNCHRONY ST
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusion: device investigation revealed the substrate of the device's circuitry to be broken.The investigation results appear to match the problems mentioned in the recipient report.The exact reason why the crack occurred cannot be determined, however a review of the dhr has been performed and no abnormality was revealed, thus the device was released according to specifications.This is a final report.
 
Event or Problem Description
The user's hearing performance with the device is affected.There was no redness or swelling seen over the implant area.Re-implantation took place on (b)(6) 2025.
 
Event or Problem Description
The user's hearing performance with the device is affected.There was no redness or swelling seen over the implant area.Re-implantation is being considered.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key22699166
Report Number9710014-2025-00704
Device Sequence Number8945006
Product Code MCM
UDI-Device Identifier09008737331825
UDI-Public(01)09008737331825
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/01/2022
Device Model NumberMI1210 SYNCHRONY ST
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Supplement Date Received by Manufacturer07/25/2025
Initial Report FDA Received Date08/04/2025
Supplement Report FDA Received Date01/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 YR
Patient SexMale
-
-