| Model Number |
DDU-E2400 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although the customer initially reported a service code and blank screen, analysis of aed determined that reported issues were due to the previous battery fully depleting within the device.Further review found that the customer's failure to promptly address red asi warnings reduced battery life expectancy.As a follow up with the customer, the proper maintenance of this device was reviewed.
|
| |
|
Event or Problem Description
|
|
A customer reported that their aed's asi is flashing red, and it 's screen is blank.They reported that this did not occur during patient use.
|
| |
|