|
Additional information was provided in h.3., h.6.And h.11.The product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint.A non qualified viscoelastic was indicated.The root cause may be related to a failure to follow the instructions for use (ifu).A non qualified viscoelastic was used in the device.The ifu instructs: during device preparation and implantation of the company iq iol with the company preloaded delivery system, an alcon qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The ifu instructs: take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the fill to line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.After the lens has been advanced to the fill to line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the fill to line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be out of position can result in a broken haptic or other negative outcome.Lens may become stuck in the device: if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become stuck in the device due to the use of a non qualified viscoelastic.Material properties of non qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (more than 23 degree c or 73 degree f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|