Please note correction to d4 (catalog #, lot #, and gtin).The reported event could be confirmed with the provided images in which signs of metallosis and bone erosion are evident.The device inspection revealed the following: all the screws and plate were visually inspected, slight discoloration in the plate and screws was observed, holes of the implant plate and hexagonal recess of the screw heads were slightly deformed indicating application of excess torsional force which also has resulted in the stripping of the threads of the screw.However, it is not feasible to determine whether this has happened during implantation or explantation of the implants.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In this case, the opinion of a medical expert was request to determine the root cause of the reported metallosis.His statement based on the information provided states: "the literature talks about the varying corrosive effects of anodized titanium, relating to the thickness and color of the anodized layers.At a fracture site, wolff¿s law is occurring with a generation of an electrical current, although very slight, the current occurs within the solutions around the plate and soft-tissue and will corrode to metal surface.It is the interaction of an inorganic material, and soft tissue and the degree of roughness of the implant surface, the methods utilized in achieving the surface topography, and the chemistry of implant materials, are some of the factors that may affect the soft tissue response of the plate to the adjacent soft tissue and create the grayish color.There has been no risk reported with this corrosive action since the levels of local metal ions and systemic ions is usually at very low levels in these cases.In case of high levels of metal ions, risks have been reported associated with local soft tissue damage, delayed hypersensitivity reaction (type iv) and systemic toxicity.As a summary, the appearance of the metallosis can be attributed to patient factors.Indeed, it is the result of how the patient reacted to the device.¿ based on hcp opinion during investigation, the root cause was attributed to a patient related issue as mentioned above.If more information is provided, the case will be reassessed.
|