ARTHREX, INC. KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN; WASHER, BOLT, NUT Back to Search Results
Model Number KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 07/15/2025, a sales representative reported via (b)(4) that (3) ar-2372blo knotless ac tightrope® open repair implants button came off the inserter before passing through the coracoid.This occurred during a case on (b)(6) 2025.There was no additional information provided, and additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6.
 
Event or Problem Description
Additional information was provided on 09/05/2025, where it was reported that this event occurred during an ac joint repair.The button fell off the inserter prematurely.The patient had a normal bone quality, that was prepped using a ar-2272.All fragments were retrieved from the patient.A dog bone button was used to complete the case.There was a minimal delay in the procedure and it was unknown if additional anesthesia was administered.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude the most likely cause.The most likely cause of the reported failure is user error, including improper bone preparation and incorrect surgical technique during device application.These factors may have contributed to the malfunction or compromised the integrity of the implant during the procedure.The complaint allegation could not be confirmed, as the device was not received for evaluation, and no evidence of the reported failure was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN
Common Device Name
WASHER, BOLT, NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22701982
Report Number1220246-2025-03223
Device Sequence Number9150412
Product Code HTN
UDI-Device Identifier00888867340152
UDI-Public(01)00888867340152(11)240801(17)290731(10)15296156
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K220947
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN
Device Catalogue NumberAR-2372BLO
Device Lot Number15296156
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date07/15/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2025
Supplement Date Received by Manufacturer09/05/2025
11/04/2025
Initial Report FDA Received Date08/04/2025
Supplement Report FDA Received Date09/23/2025
11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-