Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Synovitis (2094); Discomfort (2330); Ambulation or Postural Difficulties (2544); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Aseptic loosening (4883)
|
| Date of Event |
09/27/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: 200-02-38 - three peg patella 38mm (b)(6) 204-04-55 - trapezoid tibial tray sz 5f/5t (b)(6) 230-03-05 - optetrak asy,cr cemented femoral, sz 5, (b)(6).Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via a legal notification, that a male patient who had a right knee implanted, underwent a revision procedure of their knee device secondary to polyethylene liner wear resulting in aseptic loosening and synovitis of the right total knee arthroplasty.Despite undergoing the revision surgeries, plaintiff experiences daily knee pain and discomfort which limit plaintiff¿s activities of daily living and recreation and impacts plaintiff¿s quality of life.As a direct, proximate and legal consequence of the defective nature of the knee device as described herein, plaintiff has suffered and continues to suffer permanent and debilitating injures and damages, including but not limited to, significant pain and discomfort; gait impairment; poor balance; difficulty walking; component part loosening; metallosis; soft tissue damage; infection; and other injuries presently undiagnosed, which all require ongoing medical care.No further information available.This is 1 of 2 reports for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h3, h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A review of complaint history determined that no further action is required as no trends were identified.The reason for the reported revision cannot be confirmed from the information provided but may be the result of prosthesis wear and loosening or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear and loosening cannot be confirmed and potential contributions of user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images radiographs, or relevant clinical information were provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|