RESPIRONICS, INC. TRILOGY 100 VENTILATOR, JAPAN - BT; VENTILATOR, CONTINUOUS, FACILITY USE
|
Back to Search Results |
|
| Model Number |
1054655 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, an operational check was performed and the (right, left) buttons reacted as if pressed twice when pressed only once.The front panel/keypad led pca was replaced to address the issue.Additionally, the front enclosure overlay was replaced due to scratches.This is not related to the reportable malfunction/issue.
|
| |
|