| Model Number |
9-ASD-008 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 8mm amplatzer septal occluder was chosen for an iatrogenic atrial septal defect (asd) post left sided procedure using an unknown delivery system.During procedure, the device had cobra headed deformation and was unable to be used.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The device was removed from the body and a new 10mm amplatzer septal occluder was chosen and completed the procedure.The patient remained stable throughout the procedure and no delays to the case occurred.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on available information and without the device to analyze, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|