Catalog Number MP5313-C
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd maxguard pressure rated extension set was leaking the following information was received by the initial reporter with the following verbatim it was reported by customer that leakage occurred reported between the insyte ag catheter hub and the j-loop extension set during a power injection in ct.
 
Event or Problem Description
No additional information was provided.
 
Additional Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint of leaking at the j-loop during power injection could not be verified due to the product not being returned for failure investigation.The mp5313-c set is rated for power injection according to the ifu under the parameters 'pressure rated up to 325 psi at a flow rate of 10 ml/second'.If the needs of the customer are outside of these parameters, please reach out to your bd sales rep for guidance.The root cause for the connection and leaking issue could not be found without a sample investigation.We understand the challenging connection between the spin collar and the male luer component to the catheter hub, and bd is available to help if the customer has any further issues.A device history record review could not be performed because the lot number is unknown.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXPLUS PRESSURE RATED EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22708611
Report Number9616066-2025-02228
Device Sequence Number13201436
Product Code FPA
UDI-Device Identifier10885403236716
UDI-Public(01)10885403236716
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K083472
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMP5313-C
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2025
Supplement Date Received by Manufacturer09/10/2025
Initial Report FDA Received Date08/04/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-